Topology optimization method applied to ventricular assis device impeller and volute design.

Detalhes bibliográficos
Ano de defesa: 2019
Autor(a) principal: Sá, Luís Fernando Nogueira de
Orientador(a): Não Informado pela instituição
Banca de defesa: Não Informado pela instituição
Tipo de documento: Tese
Tipo de acesso: Acesso aberto
Idioma: eng
Instituição de defesa: Biblioteca Digitais de Teses e Dissertações da USP
Programa de Pós-Graduação: Não Informado pela instituição
Departamento: Não Informado pela instituição
País: Não Informado pela instituição
Palavras-chave em Português:
VAD
Link de acesso: http://www.teses.usp.br/teses/disponiveis/3/3152/tde-06022020-110012/
Resumo: Ventricular assist devices (VADs) are mainly small-scale pumps that assist the blood to flow. The design of VAD has as the primary requirement low shear stress in order to avoid the death of red blood cells. The topologies of the rotor and volute of the VAD can directly influence the flow condition. This work aims to optimize the design of the rotor and volute of a radial VAD, based on the topological optimization method. The constitutive equations are solved by using the finite element method for the Carreau-Yasuda blood flow model, in a rotational reference in the case of the rotor, and in a fixed reference frame in the case of the volute. The fluid flow is modeled in porous media using the Brinkman equation; also, the turbulence model of Spalart-Allmaras is used. The objective function is the minimization of viscous energy dissipation, aiming to minimize the hemolysis rate indirectly. The implementation of the algorithms is performed using the FEniCS environment coupled with the dolfin-adjoint and pyIpopt libraries. The optimization problem is solved using the PyIpopt algorithm. The optimized topologies are built using 3D printing. The prototypes are tested with water, and the results are used to calibrate a 3D complete VAD computational model. Finally, the model is simulated with blood at 5l/min and 100mmHg and compared with a straight blade impeller. A reduction of 20% in the viscous dissipation is observed.