Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection
Main Author: | |
---|---|
Publication Date: | 2005 |
Other Authors: | , , , , , |
Format: | Article |
Language: | eng |
Source: | Repositório Institucional da UNIFESP |
Download full: | http://dx.doi.org/10.1111/j.1432-2277.2004.00066.x http://repositorio.unifesp.br/handle/11600/28163 |
Summary: | The need to adapt optimal conditions of sirolimus blood level monitoring in laboratories led us to optimize an high-performance liquid chromatography-ultraviolet method and compare the elution performances using the mobile phase A, 68% MeOH/2% acetonitrile (ACN)/30% H2O and mobile phase B, 30% MeOH/42% ACN/28% H2O. Samples were assayed with 1-chlorobutane, redissolved in MeOH/water and injected onto a C-18 column at 50 degrees C. the assay achieved sensitivity of 2.5-150 ng/ml (CV = 10.6%) and recovery of 92-103.6%. the intra- and interassay precisions ranged from 3.3% to 13% and from 5.9% to 15% for quality controls of 7.5, 60 and 120 ng/ml. the mobile phase A was unable to elute and recover sirolimus and internal standard in the expected retention time and concentration. Under our working conditions, the assay was precise, accurate and sensible, stressing the importance of establishing for the best working conditions according to the staff and demands of the laboratory. |
id |
UFSP_e962d76d74b51ef9f33b4d185637f702 |
---|---|
oai_identifier_str |
oai:repositorio.unifesp.br/:11600/28163 |
network_acronym_str |
UFSP |
network_name_str |
Repositório Institucional da UNIFESP |
repository_id_str |
3465 |
spelling |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detectionhigh-performance liquid chromatographysirolimusultraviolet detectionwhole bloodThe need to adapt optimal conditions of sirolimus blood level monitoring in laboratories led us to optimize an high-performance liquid chromatography-ultraviolet method and compare the elution performances using the mobile phase A, 68% MeOH/2% acetonitrile (ACN)/30% H2O and mobile phase B, 30% MeOH/42% ACN/28% H2O. Samples were assayed with 1-chlorobutane, redissolved in MeOH/water and injected onto a C-18 column at 50 degrees C. the assay achieved sensitivity of 2.5-150 ng/ml (CV = 10.6%) and recovery of 92-103.6%. the intra- and interassay precisions ranged from 3.3% to 13% and from 5.9% to 15% for quality controls of 7.5, 60 and 120 ng/ml. the mobile phase A was unable to elute and recover sirolimus and internal standard in the expected retention time and concentration. Under our working conditions, the assay was precise, accurate and sensible, stressing the importance of establishing for the best working conditions according to the staff and demands of the laboratory.Universidade Federal de São Paulo, Escola Paulista Med, Dept Med, Div Nephrol, BR-04023900 São Paulo, BrazilUniversidade Federal de São Paulo, Fdn Oswaldo Ramos, Dept Med, Div Nephrol, BR-04023900 São Paulo, BrazilWyeth Ayerst Res, São Paulo, BrazilUniversidade Federal de São Paulo, Escola Paulista Med, Dept Med, Div Nephrol, BR-04023900 São Paulo, BrazilUniversidade Federal de São Paulo, Fdn Oswaldo Ramos, Dept Med, Div Nephrol, BR-04023900 São Paulo, BrazilWeb of ScienceBlackwell Publishing LtdUniversidade Federal de São Paulo (UNIFESP)Wyeth Ayerst ResAndrade, Maria Claudina Camargo de [UNIFESP]Di Marco, Giovana Seno [UNIFESP]Felipe, Claudia Rosso [UNIFESP]Alfieri, FernandoTedesco-Silva Junior, Hélio [UNIFESP]Pestana, José Osmar Medina [UNIFESP]Casarini, Dulce Elena [UNIFESP]2016-01-24T12:37:41Z2016-01-24T12:37:41Z2005-03-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersion354-359http://dx.doi.org/10.1111/j.1432-2277.2004.00066.xTransplant International. Oxford: Blackwell Publishing Ltd, v. 18, n. 3, p. 354-359, 2005.10.1111/j.1432-2277.2004.00066.x0934-0874http://repositorio.unifesp.br/handle/11600/28163WOS:000227714800014engTransplant Internationalinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UNIFESPinstname:Universidade Federal de São Paulo (UNIFESP)instacron:UNIFESP2016-01-24T10:37:41Zoai:repositorio.unifesp.br/:11600/28163Repositório InstitucionalPUBhttp://www.repositorio.unifesp.br/oai/requestbiblioteca.csp@unifesp.bropendoar:34652016-01-24T10:37:41Repositório Institucional da UNIFESP - Universidade Federal de São Paulo (UNIFESP)false |
dc.title.none.fl_str_mv |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
title |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
spellingShingle |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection Andrade, Maria Claudina Camargo de [UNIFESP] high-performance liquid chromatography sirolimus ultraviolet detection whole blood |
title_short |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
title_full |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
title_fullStr |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
title_full_unstemmed |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
title_sort |
Sirolimus quantification by high-performance liquid chromatography with ultraviolet detection |
author |
Andrade, Maria Claudina Camargo de [UNIFESP] |
author_facet |
Andrade, Maria Claudina Camargo de [UNIFESP] Di Marco, Giovana Seno [UNIFESP] Felipe, Claudia Rosso [UNIFESP] Alfieri, Fernando Tedesco-Silva Junior, Hélio [UNIFESP] Pestana, José Osmar Medina [UNIFESP] Casarini, Dulce Elena [UNIFESP] |
author_role |
author |
author2 |
Di Marco, Giovana Seno [UNIFESP] Felipe, Claudia Rosso [UNIFESP] Alfieri, Fernando Tedesco-Silva Junior, Hélio [UNIFESP] Pestana, José Osmar Medina [UNIFESP] Casarini, Dulce Elena [UNIFESP] |
author2_role |
author author author author author author |
dc.contributor.none.fl_str_mv |
Universidade Federal de São Paulo (UNIFESP) Wyeth Ayerst Res |
dc.contributor.author.fl_str_mv |
Andrade, Maria Claudina Camargo de [UNIFESP] Di Marco, Giovana Seno [UNIFESP] Felipe, Claudia Rosso [UNIFESP] Alfieri, Fernando Tedesco-Silva Junior, Hélio [UNIFESP] Pestana, José Osmar Medina [UNIFESP] Casarini, Dulce Elena [UNIFESP] |
dc.subject.por.fl_str_mv |
high-performance liquid chromatography sirolimus ultraviolet detection whole blood |
topic |
high-performance liquid chromatography sirolimus ultraviolet detection whole blood |
description |
The need to adapt optimal conditions of sirolimus blood level monitoring in laboratories led us to optimize an high-performance liquid chromatography-ultraviolet method and compare the elution performances using the mobile phase A, 68% MeOH/2% acetonitrile (ACN)/30% H2O and mobile phase B, 30% MeOH/42% ACN/28% H2O. Samples were assayed with 1-chlorobutane, redissolved in MeOH/water and injected onto a C-18 column at 50 degrees C. the assay achieved sensitivity of 2.5-150 ng/ml (CV = 10.6%) and recovery of 92-103.6%. the intra- and interassay precisions ranged from 3.3% to 13% and from 5.9% to 15% for quality controls of 7.5, 60 and 120 ng/ml. the mobile phase A was unable to elute and recover sirolimus and internal standard in the expected retention time and concentration. Under our working conditions, the assay was precise, accurate and sensible, stressing the importance of establishing for the best working conditions according to the staff and demands of the laboratory. |
publishDate |
2005 |
dc.date.none.fl_str_mv |
2005-03-01 2016-01-24T12:37:41Z 2016-01-24T12:37:41Z |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://dx.doi.org/10.1111/j.1432-2277.2004.00066.x Transplant International. Oxford: Blackwell Publishing Ltd, v. 18, n. 3, p. 354-359, 2005. 10.1111/j.1432-2277.2004.00066.x 0934-0874 http://repositorio.unifesp.br/handle/11600/28163 WOS:000227714800014 |
url |
http://dx.doi.org/10.1111/j.1432-2277.2004.00066.x http://repositorio.unifesp.br/handle/11600/28163 |
identifier_str_mv |
Transplant International. Oxford: Blackwell Publishing Ltd, v. 18, n. 3, p. 354-359, 2005. 10.1111/j.1432-2277.2004.00066.x 0934-0874 WOS:000227714800014 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
Transplant International |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
354-359 |
dc.publisher.none.fl_str_mv |
Blackwell Publishing Ltd |
publisher.none.fl_str_mv |
Blackwell Publishing Ltd |
dc.source.none.fl_str_mv |
reponame:Repositório Institucional da UNIFESP instname:Universidade Federal de São Paulo (UNIFESP) instacron:UNIFESP |
instname_str |
Universidade Federal de São Paulo (UNIFESP) |
instacron_str |
UNIFESP |
institution |
UNIFESP |
reponame_str |
Repositório Institucional da UNIFESP |
collection |
Repositório Institucional da UNIFESP |
repository.name.fl_str_mv |
Repositório Institucional da UNIFESP - Universidade Federal de São Paulo (UNIFESP) |
repository.mail.fl_str_mv |
biblioteca.csp@unifesp.br |
_version_ |
1841453731031285760 |