Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia

Bibliographic Details
Main Author: Borges, Melania Aparecida [UNIFESP]
Publication Date: 2018
Format: Master thesis
Language: por
Source: Repositório Institucional da UNIFESP
Download full: https://sucupira.capes.gov.br/sucupira/public/consultas/coleta/trabalhoConclusao/viewTrabalhoConclusao.jsf?popup=true&id_trabalho=6527077
https://repositorio.unifesp.br/handle/11600/52124
Summary: Developing and testing a blood pressure beattobeat monitoring device during coronary angiography and subsequently compare with the measurements obtained from invasive blood pressure. Methods: Twentyeight patients with an indication of hemodynamic study were included and remained in the horizontal dorsal decubitus position. Before onset coronary angiography they were guided about the use of the left radial bracelet to monitor blood pressure. Results: There was a significant difference between the time required for the hemodynamic team to acquire the first measurement of invasive blood pressure and the time of the first measurement of heart beat (11.1 ± 5.1 and 1.5 ± 1.8 p <0.0001, respectively). The intraclass correlation coefficients (95% CI) of systolic and diastolic blood pressure were 0.897 (0.780 0.952) and 0.876 (0.734 0.942), respectively, indicating good reproducibility. Conclusion: this study showed the process of developing an equipment to evaluate beattobeat blood pressure but when compared to invasive blood pressure there were no significant differences between the two measurements. This technique may be a promising coadjuvant tool associated with invasive monitoring during coronary angiography procedures.
id UFSP_a95cc3c8254deb7f3b92485dba48adaa
oai_identifier_str oai:repositorio.unifesp.br/:11600/52124
network_acronym_str UFSP
network_name_str Repositório Institucional da UNIFESP
repository_id_str 3465
spelling Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografiaDevelopment and clinical validation of equipment for blood pressure measurement noninvasive heart beat compared to invasive blood pressure during coronary angiographyCoronary angiographyBlood pressureValidationMonitoring, physiologicAngiografia coronáriaPressão sanguíneaValidaçãoMonitorização fisiológicaDeveloping and testing a blood pressure beattobeat monitoring device during coronary angiography and subsequently compare with the measurements obtained from invasive blood pressure. Methods: Twentyeight patients with an indication of hemodynamic study were included and remained in the horizontal dorsal decubitus position. Before onset coronary angiography they were guided about the use of the left radial bracelet to monitor blood pressure. Results: There was a significant difference between the time required for the hemodynamic team to acquire the first measurement of invasive blood pressure and the time of the first measurement of heart beat (11.1 ± 5.1 and 1.5 ± 1.8 p <0.0001, respectively). The intraclass correlation coefficients (95% CI) of systolic and diastolic blood pressure were 0.897 (0.780 0.952) and 0.876 (0.734 0.942), respectively, indicating good reproducibility. Conclusion: this study showed the process of developing an equipment to evaluate beattobeat blood pressure but when compared to invasive blood pressure there were no significant differences between the two measurements. This technique may be a promising coadjuvant tool associated with invasive monitoring during coronary angiography procedures.Fundamento: Os equipamentos de avaliação da pressão arterial batimento a batimento apresentam boa acurácia quando comparado com a pressão arterial invasiva e podem contribuir com a monitorização hemodinâmica durante a realização de procedimentos intervencionistas. Entretanto, seu uso na prática clínica ainda não foi totalmente validado. Objetivo: Desenvolver e testar um equipamento de monitorização de pressão arterial batimento a batimento durante a realização de coronariografia e comparar com as medidas obtidas pela pressão arterial invasiva. Métodos: Foram selecionados 28 pacientes com indicação de estudo hemodinâmico. Os pacientes permaneceram em decúbito dorsal horizontal e foram orientados quanto ao uso da pulseira no pulso radial esquerdo para monitorização da pressão arterial periférica batimento a batimento. Após a realização da punção da artéria femural ou radial direita e verificação da pressão arterial invasiva através de transdutor de pressão foi realizada a calibração do sistema de tonometria arterial periférica. As medidas pressóricas foram coletadas em 6 momentos durante a coronariografia em intervalos regulares e os dados foram comparados com as medidas simultâneas obtidas pelo sistema invasivo. Todos os pacientes assinaram o termo de consentimento livre e esclarecido. A comparação de médias das duas mensurações foi realizada utilizandose teste t de Student para amostras pareadas. Valor de p ≤ 0,05 foi considerado significante. CEP: 534.636 Resultados: Houve diferença significativa entre o tempo necessário para a equipe de hemodinâmica adquirir a primeira medida da pressão arterial invasiva e o tempo da primeira medida da pressão arterial batimento a batimento (11,1 ± 5,1 e 1,5 ± 1,8; p < 0,0001, respectivamente). Não houve diferença estatística entre as médias da pressão arterial sistólica com o equipamento quando comparado com a pressão arterial invasiva (130,73 ± 15,95 vs 128,79 ± 16,24, p= 0,305, respectivamente). Não houve diferença estatística entre as médias da pressão arterial diastólica com o equipamento quando comparado com a pressão arterial invasiva (78,26 ± 13,23 vs 77,57 ± 12,54, p= 0,677, respectivamente). Os coeficientes de correlação intraclasse (IC 95%) da pressão arterial sistólica e da diastólica foram 0,897 (0,780; 0,952) e 0,876 (0,734; 0,942), respectivamente, indicando uma boa reprodutibilidade. Conclusão: O equipamento desenvolvido foi testado e não houve diferenças significativas entre as medidas de pressão arterial invasiva e por tonometria arterial periférica. Assim, esta técnica pode constituir uma ferramenta coadjuvante promissora, quando associada à monitorização invasiva, durante procedimentos de coronariografia.Dados abertos - Sucupira - Teses e dissertações (2018)Coordenação de Aperfeiçoamento de Pessoal de Nível Superior (CAPES)Universidade Federal de São Paulo (UNIFESP)Cintra, Fatima Dumas [UNIFESP]http://lattes.cnpq.br/8836659601497424http://lattes.cnpq.br/1492283413049314Universidade Federal de São Paulo (UNIFESP)Borges, Melania Aparecida [UNIFESP]2020-03-25T11:43:21Z2020-03-25T11:43:21Z2018-11-28info:eu-repo/semantics/masterThesisinfo:eu-repo/semantics/publishedVersion55 f.application/pdfhttps://sucupira.capes.gov.br/sucupira/public/consultas/coleta/trabalhoConclusao/viewTrabalhoConclusao.jsf?popup=true&id_trabalho=65270772018-0037.pdfhttps://repositorio.unifesp.br/handle/11600/52124porSão Pauloinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UNIFESPinstname:Universidade Federal de São Paulo (UNIFESP)instacron:UNIFESP2024-08-10T06:53:30Zoai:repositorio.unifesp.br/:11600/52124Repositório InstitucionalPUBhttp://www.repositorio.unifesp.br/oai/requestbiblioteca.csp@unifesp.bropendoar:34652024-08-10T06:53:30Repositório Institucional da UNIFESP - Universidade Federal de São Paulo (UNIFESP)false
dc.title.none.fl_str_mv Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
Development and clinical validation of equipment for blood pressure measurement noninvasive heart beat compared to invasive blood pressure during coronary angiography
title Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
spellingShingle Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
Borges, Melania Aparecida [UNIFESP]
Coronary angiography
Blood pressure
Validation
Monitoring, physiologic
Angiografia coronária
Pressão sanguínea
Validação
Monitorização fisiológica
title_short Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
title_full Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
title_fullStr Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
title_full_unstemmed Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
title_sort Desenvolvimento de equipamento para medida de pressão arterial batimento a batimento comparado com a pressão arterial invasiva durante procedimentos de angiocoronariografia
author Borges, Melania Aparecida [UNIFESP]
author_facet Borges, Melania Aparecida [UNIFESP]
author_role author
dc.contributor.none.fl_str_mv Cintra, Fatima Dumas [UNIFESP]
http://lattes.cnpq.br/8836659601497424
http://lattes.cnpq.br/1492283413049314
Universidade Federal de São Paulo (UNIFESP)
dc.contributor.author.fl_str_mv Borges, Melania Aparecida [UNIFESP]
dc.subject.por.fl_str_mv Coronary angiography
Blood pressure
Validation
Monitoring, physiologic
Angiografia coronária
Pressão sanguínea
Validação
Monitorização fisiológica
topic Coronary angiography
Blood pressure
Validation
Monitoring, physiologic
Angiografia coronária
Pressão sanguínea
Validação
Monitorização fisiológica
description Developing and testing a blood pressure beattobeat monitoring device during coronary angiography and subsequently compare with the measurements obtained from invasive blood pressure. Methods: Twentyeight patients with an indication of hemodynamic study were included and remained in the horizontal dorsal decubitus position. Before onset coronary angiography they were guided about the use of the left radial bracelet to monitor blood pressure. Results: There was a significant difference between the time required for the hemodynamic team to acquire the first measurement of invasive blood pressure and the time of the first measurement of heart beat (11.1 ± 5.1 and 1.5 ± 1.8 p <0.0001, respectively). The intraclass correlation coefficients (95% CI) of systolic and diastolic blood pressure were 0.897 (0.780 0.952) and 0.876 (0.734 0.942), respectively, indicating good reproducibility. Conclusion: this study showed the process of developing an equipment to evaluate beattobeat blood pressure but when compared to invasive blood pressure there were no significant differences between the two measurements. This technique may be a promising coadjuvant tool associated with invasive monitoring during coronary angiography procedures.
publishDate 2018
dc.date.none.fl_str_mv 2018-11-28
2020-03-25T11:43:21Z
2020-03-25T11:43:21Z
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format masterThesis
status_str publishedVersion
dc.identifier.uri.fl_str_mv https://sucupira.capes.gov.br/sucupira/public/consultas/coleta/trabalhoConclusao/viewTrabalhoConclusao.jsf?popup=true&id_trabalho=6527077
2018-0037.pdf
https://repositorio.unifesp.br/handle/11600/52124
url https://sucupira.capes.gov.br/sucupira/public/consultas/coleta/trabalhoConclusao/viewTrabalhoConclusao.jsf?popup=true&id_trabalho=6527077
https://repositorio.unifesp.br/handle/11600/52124
identifier_str_mv 2018-0037.pdf
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv 55 f.
application/pdf
dc.coverage.none.fl_str_mv São Paulo
dc.publisher.none.fl_str_mv Universidade Federal de São Paulo (UNIFESP)
publisher.none.fl_str_mv Universidade Federal de São Paulo (UNIFESP)
dc.source.none.fl_str_mv reponame:Repositório Institucional da UNIFESP
instname:Universidade Federal de São Paulo (UNIFESP)
instacron:UNIFESP
instname_str Universidade Federal de São Paulo (UNIFESP)
instacron_str UNIFESP
institution UNIFESP
reponame_str Repositório Institucional da UNIFESP
collection Repositório Institucional da UNIFESP
repository.name.fl_str_mv Repositório Institucional da UNIFESP - Universidade Federal de São Paulo (UNIFESP)
repository.mail.fl_str_mv biblioteca.csp@unifesp.br
_version_ 1841453750420504576