Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study

Bibliographic Details
Main Author: Ferreira,Ana Leitão
Publication Date: 2019
Other Authors: Mendes,Joaquim Gabriel, Nunes,Catarina Sofia, Amorim,Pedro
Format: Article
Language: eng
Source: Revista Brasileira de Anestesiologia (Online)
Download full: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942019000400007
Summary: Abstract Background and objectives According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results The difference between the predicted and the real BIS was in average 30.09 ± 18.73 s. Conclusion Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.
id SBA-1_89c42b80d75444bc4e297f8fbe2164c1
oai_identifier_str oai:scielo:S0034-70942019000400007
network_acronym_str SBA-1
network_name_str Revista Brasileira de Anestesiologia (Online)
repository_id_str
spelling Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational studyBispectral index monitorDepth of anesthesiaAwarenessTime delayAbstract Background and objectives According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results The difference between the predicted and the real BIS was in average 30.09 ± 18.73 s. Conclusion Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.Sociedade Brasileira de Anestesiologia2019-08-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942019000400007Revista Brasileira de Anestesiologia v.69 n.4 2019reponame:Revista Brasileira de Anestesiologia (Online)instname:Sociedade Brasileira de Anestesiologia (SBA)instacron:SBA10.1016/j.bjane.2019.04.006info:eu-repo/semantics/openAccessFerreira,Ana LeitãoMendes,Joaquim GabrielNunes,Catarina SofiaAmorim,Pedroeng2019-10-08T00:00:00Zoai:scielo:S0034-70942019000400007Revistahttps://www.sbahq.org/revista/https://old.scielo.br/oai/scielo-oai.php||sba2000@openlink.com.br1806-907X0034-7094opendoar:2019-10-08T00:00Revista Brasileira de Anestesiologia (Online) - Sociedade Brasileira de Anestesiologia (SBA)false
dc.title.none.fl_str_mv Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
title Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
spellingShingle Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
Ferreira,Ana Leitão
Bispectral index monitor
Depth of anesthesia
Awareness
Time delay
title_short Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
title_full Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
title_fullStr Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
title_full_unstemmed Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
title_sort Evaluation of Bispectral Index time delay in response to anesthesia induction: an observational study
author Ferreira,Ana Leitão
author_facet Ferreira,Ana Leitão
Mendes,Joaquim Gabriel
Nunes,Catarina Sofia
Amorim,Pedro
author_role author
author2 Mendes,Joaquim Gabriel
Nunes,Catarina Sofia
Amorim,Pedro
author2_role author
author
author
dc.contributor.author.fl_str_mv Ferreira,Ana Leitão
Mendes,Joaquim Gabriel
Nunes,Catarina Sofia
Amorim,Pedro
dc.subject.por.fl_str_mv Bispectral index monitor
Depth of anesthesia
Awareness
Time delay
topic Bispectral index monitor
Depth of anesthesia
Awareness
Time delay
description Abstract Background and objectives According to the manufacturer, the Bispectral Index (BIS) has a processing time delay of 5-10 s. Studies addressing this have suggested longer delays. We evaluated the time delay in the Bispectral Index response. Methods Based on clinical data from 45 patients, using the difference between the predicted and the real BIS, calculated during a fixed 3 minutes period after the moment the Bispectral Index dropped below 80 during the induction of general anesthesia with propofol and remifentanil. Results The difference between the predicted and the real BIS was in average 30.09 ± 18.73 s. Conclusion Our results may be another indication that the delay in BIS processing may be much longer than stated by the manufacture, a fact with clinical implications.
publishDate 2019
dc.date.none.fl_str_mv 2019-08-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942019000400007
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0034-70942019000400007
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1016/j.bjane.2019.04.006
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Sociedade Brasileira de Anestesiologia
publisher.none.fl_str_mv Sociedade Brasileira de Anestesiologia
dc.source.none.fl_str_mv Revista Brasileira de Anestesiologia v.69 n.4 2019
reponame:Revista Brasileira de Anestesiologia (Online)
instname:Sociedade Brasileira de Anestesiologia (SBA)
instacron:SBA
instname_str Sociedade Brasileira de Anestesiologia (SBA)
instacron_str SBA
institution SBA
reponame_str Revista Brasileira de Anestesiologia (Online)
collection Revista Brasileira de Anestesiologia (Online)
repository.name.fl_str_mv Revista Brasileira de Anestesiologia (Online) - Sociedade Brasileira de Anestesiologia (SBA)
repository.mail.fl_str_mv ||sba2000@openlink.com.br
_version_ 1752126630478741504