Summary: Ventricular oversensing in patients with defibrillators is an infrequent but deleterious condition. We report a patient with a cardiac resynchronization-defibrillation device that presented with hyperkalemia and syncope. Device interrogation revealed ventricular double-counting within the QRS of a slow ventricular tachycardia, resulting detection of the slow ventricular tachycardia in the ventricular fibrillation zone, and delivery of an effective therapy, below device programmed detection rate. This case of defibrillator inappropriate detection emphasizes the relevance of device electrogram interrogation in order to minimize inappropriate therapies.
id RCAP_f5648633c9d54b78596dd24a34182ce7
oai_identifier_str oai:repositorio.hff.min-saude.pt:10400.10/1586
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling An inappropriate defibrillator shock during ventricular tachycardia.Implantable defibrillatorsElectrocardiographyHyperkalemiaVentricular tachycardiaDesfibrilhadores implantáveisTaquicardia ventricularVentricular oversensing in patients with defibrillators is an infrequent but deleterious condition. We report a patient with a cardiac resynchronization-defibrillation device that presented with hyperkalemia and syncope. Device interrogation revealed ventricular double-counting within the QRS of a slow ventricular tachycardia, resulting detection of the slow ventricular tachycardia in the ventricular fibrillation zone, and delivery of an effective therapy, below device programmed detection rate. This case of defibrillator inappropriate detection emphasizes the relevance of device electrogram interrogation in order to minimize inappropriate therapies.Indian Heart Rhythm SocietyUnidade Local de Saúde Amadora / SintraFaustino, MMadeira, FMorais, C2016-03-28T11:23:01Z20152015-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.10/1586eng0972-629210.1016/j.ipej.2015.10.002info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-03-10T15:02:36Zoai:repositorio.hff.min-saude.pt:10400.10/1586Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-29T01:15:42.017046Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv An inappropriate defibrillator shock during ventricular tachycardia.
title An inappropriate defibrillator shock during ventricular tachycardia.
spellingShingle An inappropriate defibrillator shock during ventricular tachycardia.
Faustino, M
Implantable defibrillators
Electrocardiography
Hyperkalemia
Ventricular tachycardia
Desfibrilhadores implantáveis
Taquicardia ventricular
title_short An inappropriate defibrillator shock during ventricular tachycardia.
title_full An inappropriate defibrillator shock during ventricular tachycardia.
title_fullStr An inappropriate defibrillator shock during ventricular tachycardia.
title_full_unstemmed An inappropriate defibrillator shock during ventricular tachycardia.
title_sort An inappropriate defibrillator shock during ventricular tachycardia.
author Faustino, M
author_facet Faustino, M
Madeira, F
Morais, C
author_role author
author2 Madeira, F
Morais, C
author2_role author
author
dc.contributor.none.fl_str_mv Unidade Local de Saúde Amadora / Sintra
dc.contributor.author.fl_str_mv Faustino, M
Madeira, F
Morais, C
dc.subject.por.fl_str_mv Implantable defibrillators
Electrocardiography
Hyperkalemia
Ventricular tachycardia
Desfibrilhadores implantáveis
Taquicardia ventricular
topic Implantable defibrillators
Electrocardiography
Hyperkalemia
Ventricular tachycardia
Desfibrilhadores implantáveis
Taquicardia ventricular
description Ventricular oversensing in patients with defibrillators is an infrequent but deleterious condition. We report a patient with a cardiac resynchronization-defibrillation device that presented with hyperkalemia and syncope. Device interrogation revealed ventricular double-counting within the QRS of a slow ventricular tachycardia, resulting detection of the slow ventricular tachycardia in the ventricular fibrillation zone, and delivery of an effective therapy, below device programmed detection rate. This case of defibrillator inappropriate detection emphasizes the relevance of device electrogram interrogation in order to minimize inappropriate therapies.
publishDate 2015
dc.date.none.fl_str_mv 2015
2015-01-01T00:00:00Z
2016-03-28T11:23:01Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.10/1586
url http://hdl.handle.net/10400.10/1586
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 0972-6292
10.1016/j.ipej.2015.10.002
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Indian Heart Rhythm Society
publisher.none.fl_str_mv Indian Heart Rhythm Society
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833600897042612224