Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure

Bibliographic Details
Main Author: Aguiar Rosa, S
Publication Date: 2017
Other Authors: Ferreira, F, de Sousa, L, Fiarresga, A, Martins, JD, Galrinho, A, Agapito, A, Fazendas, P, Pinto, MF, Cruz Ferreira, R
Format: Article
Language: eng
Source: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Download full: http://hdl.handle.net/10400.17/2935
Summary: A 39-year-old woman underwent uneventful percutaneous occlusion of an ostium secundum atrial septal defect (ASD) with a 22 mm Ultrasept ASD Occluder®. Transesophageal echocardiography (TEE) performed two years after implantation revealed a de novo residual left-to-right shunt through the correctly implanted device. Three-dimensional transesophageal echocardiography (3D TEE) further clarified this finding by showing a perforation of the device membrane coating. The patient underwent transcatheter closure of the residual shunt with a 20 mm Ultrasept PFO®device. The procedure was guided by fluoroscopy and real-time 3D TEE. At the end of the procedure 3D TEE documented correct device deployment with complete defect coverage and absence of residual shunt.
id RCAP_cbf17e92b4268b35c3e7162439109b0a
oai_identifier_str oai:repositorio.chlc.pt:10400.17/2935
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device FailureEncerramento Percutâneo de Comunicação Interauricular Residual por Falência de DispositivoHSM CARHSM CAR PEDCardiac Surgical Procedures/methodsEquipment FailureHeart Septal Defects, Atrial/surgeryRemission InductionSeptal Occluder DeviceA 39-year-old woman underwent uneventful percutaneous occlusion of an ostium secundum atrial septal defect (ASD) with a 22 mm Ultrasept ASD Occluder®. Transesophageal echocardiography (TEE) performed two years after implantation revealed a de novo residual left-to-right shunt through the correctly implanted device. Three-dimensional transesophageal echocardiography (3D TEE) further clarified this finding by showing a perforation of the device membrane coating. The patient underwent transcatheter closure of the residual shunt with a 20 mm Ultrasept PFO®device. The procedure was guided by fluoroscopy and real-time 3D TEE. At the end of the procedure 3D TEE documented correct device deployment with complete defect coverage and absence of residual shunt.Sociedade Portuguesa de CardiologiaRepositório da Unidade Local de Saúde São JoséAguiar Rosa, SFerreira, Fde Sousa, LFiarresga, AMartins, JDGalrinho, AAgapito, AFazendas, PPinto, MFCruz Ferreira, R2018-03-06T16:40:27Z2017-062017-06-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.17/2935eng10.1016/j.repc.2016.09.020info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-03-06T16:51:26Zoai:repositorio.chlc.pt:10400.17/2935Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-29T00:22:05.122454Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
Encerramento Percutâneo de Comunicação Interauricular Residual por Falência de Dispositivo
title Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
spellingShingle Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
Aguiar Rosa, S
HSM CAR
HSM CAR PED
Cardiac Surgical Procedures/methods
Equipment Failure
Heart Septal Defects, Atrial/surgery
Remission Induction
Septal Occluder Device
title_short Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
title_full Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
title_fullStr Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
title_full_unstemmed Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
title_sort Successful Percutaneous Closure of a Residual Atrial Septal Defect Due to Device Failure
author Aguiar Rosa, S
author_facet Aguiar Rosa, S
Ferreira, F
de Sousa, L
Fiarresga, A
Martins, JD
Galrinho, A
Agapito, A
Fazendas, P
Pinto, MF
Cruz Ferreira, R
author_role author
author2 Ferreira, F
de Sousa, L
Fiarresga, A
Martins, JD
Galrinho, A
Agapito, A
Fazendas, P
Pinto, MF
Cruz Ferreira, R
author2_role author
author
author
author
author
author
author
author
author
dc.contributor.none.fl_str_mv Repositório da Unidade Local de Saúde São José
dc.contributor.author.fl_str_mv Aguiar Rosa, S
Ferreira, F
de Sousa, L
Fiarresga, A
Martins, JD
Galrinho, A
Agapito, A
Fazendas, P
Pinto, MF
Cruz Ferreira, R
dc.subject.por.fl_str_mv HSM CAR
HSM CAR PED
Cardiac Surgical Procedures/methods
Equipment Failure
Heart Septal Defects, Atrial/surgery
Remission Induction
Septal Occluder Device
topic HSM CAR
HSM CAR PED
Cardiac Surgical Procedures/methods
Equipment Failure
Heart Septal Defects, Atrial/surgery
Remission Induction
Septal Occluder Device
description A 39-year-old woman underwent uneventful percutaneous occlusion of an ostium secundum atrial septal defect (ASD) with a 22 mm Ultrasept ASD Occluder®. Transesophageal echocardiography (TEE) performed two years after implantation revealed a de novo residual left-to-right shunt through the correctly implanted device. Three-dimensional transesophageal echocardiography (3D TEE) further clarified this finding by showing a perforation of the device membrane coating. The patient underwent transcatheter closure of the residual shunt with a 20 mm Ultrasept PFO®device. The procedure was guided by fluoroscopy and real-time 3D TEE. At the end of the procedure 3D TEE documented correct device deployment with complete defect coverage and absence of residual shunt.
publishDate 2017
dc.date.none.fl_str_mv 2017-06
2017-06-01T00:00:00Z
2018-03-06T16:40:27Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.17/2935
url http://hdl.handle.net/10400.17/2935
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.1016/j.repc.2016.09.020
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Sociedade Portuguesa de Cardiologia
publisher.none.fl_str_mv Sociedade Portuguesa de Cardiologia
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833600506691321856