A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer

Bibliographic Details
Main Author: Jorge, Jorge
Publication Date: 2011
Other Authors: González-Méijome, José Manuel, Queirós, A., Fernandes, Paulo Rodrigues, Diaz-Rey, José Alberto
Format: Article
Language: eng
Source: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Download full: http://hdl.handle.net/1822/17144
Summary: P>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye.
id RCAP_bc1b745a320ffc6959e3951ca89dc01e
oai_identifier_str oai:repositorium.sdum.uminho.pt:1822/17144
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzerNon-contact tonometryReichert R7Goldmann applanation tonometryGoldmann applanation tonometernon-contact tonometerIntraocular pressureOcular response analyzerScience & TechnologyP>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye.WileyUniversidade do MinhoJorge, JorgeGonzález-Méijome, José ManuelQueirós, A.Fernandes, Paulo RodriguesDiaz-Rey, José Alberto20112011-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/17144eng1475-131310.1111/j.1475-1313.2010.00817.x21309804http://onlinelibrary.wiley.com/info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2024-05-11T04:30:11Zoai:repositorium.sdum.uminho.pt:1822/17144Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-28T14:50:05.195565Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
title A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
spellingShingle A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
Jorge, Jorge
Non-contact tonometry
Reichert R7
Goldmann applanation tonometry
Goldmann applanation tonometer
non-contact tonometer
Intraocular pressure
Ocular response analyzer
Science & Technology
title_short A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
title_full A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
title_fullStr A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
title_full_unstemmed A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
title_sort A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
author Jorge, Jorge
author_facet Jorge, Jorge
González-Méijome, José Manuel
Queirós, A.
Fernandes, Paulo Rodrigues
Diaz-Rey, José Alberto
author_role author
author2 González-Méijome, José Manuel
Queirós, A.
Fernandes, Paulo Rodrigues
Diaz-Rey, José Alberto
author2_role author
author
author
author
dc.contributor.none.fl_str_mv Universidade do Minho
dc.contributor.author.fl_str_mv Jorge, Jorge
González-Méijome, José Manuel
Queirós, A.
Fernandes, Paulo Rodrigues
Diaz-Rey, José Alberto
dc.subject.por.fl_str_mv Non-contact tonometry
Reichert R7
Goldmann applanation tonometry
Goldmann applanation tonometer
non-contact tonometer
Intraocular pressure
Ocular response analyzer
Science & Technology
topic Non-contact tonometry
Reichert R7
Goldmann applanation tonometry
Goldmann applanation tonometer
non-contact tonometer
Intraocular pressure
Ocular response analyzer
Science & Technology
description P>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye.
publishDate 2011
dc.date.none.fl_str_mv 2011
2011-01-01T00:00:00Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/1822/17144
url http://hdl.handle.net/1822/17144
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 1475-1313
10.1111/j.1475-1313.2010.00817.x
21309804
http://onlinelibrary.wiley.com/
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Wiley
publisher.none.fl_str_mv Wiley
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833594919775633408