A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer
Main Author: | |
---|---|
Publication Date: | 2011 |
Other Authors: | , , , |
Format: | Article |
Language: | eng |
Source: | Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
Download full: | http://hdl.handle.net/1822/17144 |
Summary: | P>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye. |
id |
RCAP_bc1b745a320ffc6959e3951ca89dc01e |
---|---|
oai_identifier_str |
oai:repositorium.sdum.uminho.pt:1822/17144 |
network_acronym_str |
RCAP |
network_name_str |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
repository_id_str |
https://opendoar.ac.uk/repository/7160 |
spelling |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzerNon-contact tonometryReichert R7Goldmann applanation tonometryGoldmann applanation tonometernon-contact tonometerIntraocular pressureOcular response analyzerScience & TechnologyP>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye.WileyUniversidade do MinhoJorge, JorgeGonzález-Méijome, José ManuelQueirós, A.Fernandes, Paulo RodriguesDiaz-Rey, José Alberto20112011-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/1822/17144eng1475-131310.1111/j.1475-1313.2010.00817.x21309804http://onlinelibrary.wiley.com/info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2024-05-11T04:30:11Zoai:repositorium.sdum.uminho.pt:1822/17144Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-28T14:50:05.195565Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse |
dc.title.none.fl_str_mv |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
title |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
spellingShingle |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer Jorge, Jorge Non-contact tonometry Reichert R7 Goldmann applanation tonometry Goldmann applanation tonometer non-contact tonometer Intraocular pressure Ocular response analyzer Science & Technology |
title_short |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
title_full |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
title_fullStr |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
title_full_unstemmed |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
title_sort |
A comparison of the NCT Reichert R7 with Goldmann applanation tonometry and the Reichert ocular response analyzer |
author |
Jorge, Jorge |
author_facet |
Jorge, Jorge González-Méijome, José Manuel Queirós, A. Fernandes, Paulo Rodrigues Diaz-Rey, José Alberto |
author_role |
author |
author2 |
González-Méijome, José Manuel Queirós, A. Fernandes, Paulo Rodrigues Diaz-Rey, José Alberto |
author2_role |
author author author author |
dc.contributor.none.fl_str_mv |
Universidade do Minho |
dc.contributor.author.fl_str_mv |
Jorge, Jorge González-Méijome, José Manuel Queirós, A. Fernandes, Paulo Rodrigues Diaz-Rey, José Alberto |
dc.subject.por.fl_str_mv |
Non-contact tonometry Reichert R7 Goldmann applanation tonometry Goldmann applanation tonometer non-contact tonometer Intraocular pressure Ocular response analyzer Science & Technology |
topic |
Non-contact tonometry Reichert R7 Goldmann applanation tonometry Goldmann applanation tonometer non-contact tonometer Intraocular pressure Ocular response analyzer Science & Technology |
description |
P>Purpose:The aim of this study was to evaluate the accuracy of intraocular pressure (IOP) values from the new non-contact tonometer (NCT) Reichert R7 by comparing results with those from Goldmann applanation tonometry (GAT) and the Reichert Ocular Response Analyzer (ORA). Other ocular dimensions were assessed to evaluate their potential influence on the IOP values obtained.Methods:Ninety two right eyes from 92 adults aged between 21 and 59 years (mean 34.9 +/- 11.7 years) were enrolled in this study. IOP was measured with R7, ORA and GAT. All measurements were taken between 14:00 and 16:00 in the afternoon. Corneal resistance factor (CRF) and corneal hysteresis were measured with the Reichert ORA. The spherical equivalent refractive error was obtained using an open field auto-refractor (WAM5500; Grand Seiko) and corneal curvature, anterior chamber depth, corneal diameter and axial length were assessed with an optical coherence biometer (IOL Master; Zeiss Meditec, CA, USA).Results:The mean values for IOP measurements were 15.20 +/- 3.37 mmHg (R7), 13.49 +/- 3.55 mmHg (GAT), 15.01 +/- 3.38 mmHg (ORA IOPcc) and 14.44 +/- 3.47 mmHg (ORA IOPg). With the exception of the CRF (rho = 0.72 p < 0.001) the correlations between ocular parameters and IOP obtained with the R7 were neither statistically nor clinically significant.Conclusions:The new NCT, R7 overestimated the IOP compared with GAT in normal, healthy eyes by about 1.7 mmHg on average (95% confidence range of approximately -2 to +6mmHg). The measures provided by the R7 were significantly influenced by the stiffness of the corneal tissue as measured by the ORA CRF value but not by other dimensional parameters of the eye. |
publishDate |
2011 |
dc.date.none.fl_str_mv |
2011 2011-01-01T00:00:00Z |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://hdl.handle.net/1822/17144 |
url |
http://hdl.handle.net/1822/17144 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
1475-1313 10.1111/j.1475-1313.2010.00817.x 21309804 http://onlinelibrary.wiley.com/ |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
application/pdf |
dc.publisher.none.fl_str_mv |
Wiley |
publisher.none.fl_str_mv |
Wiley |
dc.source.none.fl_str_mv |
reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia instacron:RCAAP |
instname_str |
FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia |
instacron_str |
RCAAP |
institution |
RCAAP |
reponame_str |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
collection |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
repository.name.fl_str_mv |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia |
repository.mail.fl_str_mv |
info@rcaap.pt |
_version_ |
1833594919775633408 |