Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
| Autor(a) principal: | |
|---|---|
| Data de Publicação: | 2022 |
| Outros Autores: | , |
| Idioma: | eng |
| Título da fonte: | Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
| Texto Completo: | http://hdl.handle.net/10400.1/18212 |
Resumo: | Dose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol. |
| id |
RCAP_a2938868ff49607ef9edcbff84be75c5 |
|---|---|
| oai_identifier_str |
oai:sapientia.ualg.pt:10400.1/18212 |
| network_acronym_str |
RCAP |
| network_name_str |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
| repository_id_str |
https://opendoar.ac.uk/repository/7160 |
| spelling |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?RadiographersCone beam CTDilationRadiation safetyTechnical aspectsDosimetric comparisonQuality assuranceDose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol.European Society of RadiologySapientiaAlmeida, RuiAbrantes, AntónioSilva, Carlos Otero2022-09-05T12:25:32Z20222022-01-01T00:00:00Zconference objectinfo:eu-repo/semantics/publishedVersionapplication/pdfhttp://hdl.handle.net/10400.1/18212enghttps://dx.doi.org/10.26044/ecr2022/C-17803info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-02-18T17:36:03Zoai:sapientia.ualg.pt:10400.1/18212Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-28T20:28:14.610822Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse |
| dc.title.none.fl_str_mv |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| title |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| spellingShingle |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? Almeida, Rui Radiographers Cone beam CT Dilation Radiation safety Technical aspects Dosimetric comparison Quality assurance |
| title_short |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| title_full |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| title_fullStr |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| title_full_unstemmed |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| title_sort |
Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands? |
| author |
Almeida, Rui |
| author_facet |
Almeida, Rui Abrantes, António Silva, Carlos Otero |
| author_role |
author |
| author2 |
Abrantes, António Silva, Carlos Otero |
| author2_role |
author author |
| dc.contributor.none.fl_str_mv |
Sapientia |
| dc.contributor.author.fl_str_mv |
Almeida, Rui Abrantes, António Silva, Carlos Otero |
| dc.subject.por.fl_str_mv |
Radiographers Cone beam CT Dilation Radiation safety Technical aspects Dosimetric comparison Quality assurance |
| topic |
Radiographers Cone beam CT Dilation Radiation safety Technical aspects Dosimetric comparison Quality assurance |
| description |
Dose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol. |
| publishDate |
2022 |
| dc.date.none.fl_str_mv |
2022-09-05T12:25:32Z 2022 2022-01-01T00:00:00Z |
| dc.type.driver.fl_str_mv |
conference object |
| dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
| status_str |
publishedVersion |
| dc.identifier.uri.fl_str_mv |
http://hdl.handle.net/10400.1/18212 |
| url |
http://hdl.handle.net/10400.1/18212 |
| dc.language.iso.fl_str_mv |
eng |
| language |
eng |
| dc.relation.none.fl_str_mv |
https://dx.doi.org/10.26044/ecr2022/C-17803 |
| dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
| eu_rights_str_mv |
openAccess |
| dc.format.none.fl_str_mv |
application/pdf |
| dc.publisher.none.fl_str_mv |
European Society of Radiology |
| publisher.none.fl_str_mv |
European Society of Radiology |
| dc.source.none.fl_str_mv |
reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia instacron:RCAAP |
| instname_str |
FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia |
| instacron_str |
RCAAP |
| institution |
RCAAP |
| reponame_str |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
| collection |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) |
| repository.name.fl_str_mv |
Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia |
| repository.mail.fl_str_mv |
info@rcaap.pt |
| _version_ |
1833598673941954560 |