Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?

Detalhes bibliográficos
Autor(a) principal: Almeida, Rui
Data de Publicação: 2022
Outros Autores: Abrantes, António, Silva, Carlos Otero
Idioma: eng
Título da fonte: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Texto Completo: http://hdl.handle.net/10400.1/18212
Resumo: Dose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol.
id RCAP_a2938868ff49607ef9edcbff84be75c5
oai_identifier_str oai:sapientia.ualg.pt:10400.1/18212
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?RadiographersCone beam CTDilationRadiation safetyTechnical aspectsDosimetric comparisonQuality assuranceDose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol.European Society of RadiologySapientiaAlmeida, RuiAbrantes, AntónioSilva, Carlos Otero2022-09-05T12:25:32Z20222022-01-01T00:00:00Zconference objectinfo:eu-repo/semantics/publishedVersionapplication/pdfhttp://hdl.handle.net/10400.1/18212enghttps://dx.doi.org/10.26044/ecr2022/C-17803info:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-02-18T17:36:03Zoai:sapientia.ualg.pt:10400.1/18212Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-28T20:28:14.610822Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
title Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
spellingShingle Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
Almeida, Rui
Radiographers
Cone beam CT
Dilation
Radiation safety
Technical aspects
Dosimetric comparison
Quality assurance
title_short Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
title_full Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
title_fullStr Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
title_full_unstemmed Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
title_sort Comparative dose measurements in dentomaxillofacial imaging procedures: are Barium and Bismuth shields useful for the protection of eyes and thyroid glands?
author Almeida, Rui
author_facet Almeida, Rui
Abrantes, António
Silva, Carlos Otero
author_role author
author2 Abrantes, António
Silva, Carlos Otero
author2_role author
author
dc.contributor.none.fl_str_mv Sapientia
dc.contributor.author.fl_str_mv Almeida, Rui
Abrantes, António
Silva, Carlos Otero
dc.subject.por.fl_str_mv Radiographers
Cone beam CT
Dilation
Radiation safety
Technical aspects
Dosimetric comparison
Quality assurance
topic Radiographers
Cone beam CT
Dilation
Radiation safety
Technical aspects
Dosimetric comparison
Quality assurance
description Dose measurements were performed at the eye lens and thyroid level using TLD dosimeters, in different configurations (with and without shields) and using standard protocols with different values of voltage (kV) and current-time product (mAs). A total of 103 exposures were performed on patients, 84 of them using a MSCT GE Lightspeed 16 equipment and the remaining 19 using a CBCT NewTom GiANO equipment. These exposures were subdivided into 22 smaller samples (19 samples with 5 exposures each, 1 sample with 4 exposures and 2 samples with 2 exposures each), in accordance with the protocols and configurations adopted. 54 were performed without the use of shields and the remaining 49 with the use of shields, namely, a bismuth shield for the thyroid, a barium shield for the eye lens and, also, a bismuth shield for the breast, which was tested only in the routine MSCT protocol.
publishDate 2022
dc.date.none.fl_str_mv 2022-09-05T12:25:32Z
2022
2022-01-01T00:00:00Z
dc.type.driver.fl_str_mv conference object
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.1/18212
url http://hdl.handle.net/10400.1/18212
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv https://dx.doi.org/10.26044/ecr2022/C-17803
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv European Society of Radiology
publisher.none.fl_str_mv European Society of Radiology
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833598673941954560