Export Ready — 

Development of polyaniline microarray electrodes for cadmium analysis

Bibliographic Details
Main Author: Oliveira, Marta
Publication Date: 2012
Other Authors: Viswanathan, Subramanian, Morais, Simone, Delerue-Matos, Cristina
Format: Article
Language: eng
Source: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Download full: http://hdl.handle.net/10400.22/2043
Summary: Disposable screen-printed electrodes (SPCE) were modified using a cosmetic product to partially block the electrode surface in order to obtain a microelectrode array. The microarrays formed were electropolymerized with aniline. Scanning electron microscopy was used to evaluate the modified and polymerized electrode surface. Electrochemical characteristics of the constructed sensor for cadmium analysis were evaluated by cyclic and square-wave voltammetry. Optimized stripping procedure in which the preconcentration of cadmium was achieved by depositing at –1.20 V (vs. Ag/AgCl) resulted in a well defined anodic peak at approximately –0.7 V at pH 4.6. The achieved limit of detection was 4 × 10−9 mol dm−3. Spray modified and polymerized microarray electrodes were successfully applied to quantify cadmium in fish sample digests.
id RCAP_4b542a9669f8ff66cbba05ebd74a002f
oai_identifier_str oai:recipp.ipp.pt:10400.22/2043
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling Development of polyaniline microarray electrodes for cadmium analysisScreen-printed electrodesMicroarray electrodesAnilineCadmiumDisposable screen-printed electrodes (SPCE) were modified using a cosmetic product to partially block the electrode surface in order to obtain a microelectrode array. The microarrays formed were electropolymerized with aniline. Scanning electron microscopy was used to evaluate the modified and polymerized electrode surface. Electrochemical characteristics of the constructed sensor for cadmium analysis were evaluated by cyclic and square-wave voltammetry. Optimized stripping procedure in which the preconcentration of cadmium was achieved by depositing at –1.20 V (vs. Ag/AgCl) resulted in a well defined anodic peak at approximately –0.7 V at pH 4.6. The achieved limit of detection was 4 × 10−9 mol dm−3. Spray modified and polymerized microarray electrodes were successfully applied to quantify cadmium in fish sample digests.Springer VerlagREPOSITÓRIO P.PORTOOliveira, MartaViswanathan, SubramanianMorais, SimoneDelerue-Matos, Cristina2013-10-01T09:29:18Z20122012-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.22/2043eng1336-907510.2478/s11696-012-0175-xinfo:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-04-02T02:52:34Zoai:recipp.ipp.pt:10400.22/2043Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-29T00:25:47.882780Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv Development of polyaniline microarray electrodes for cadmium analysis
title Development of polyaniline microarray electrodes for cadmium analysis
spellingShingle Development of polyaniline microarray electrodes for cadmium analysis
Oliveira, Marta
Screen-printed electrodes
Microarray electrodes
Aniline
Cadmium
title_short Development of polyaniline microarray electrodes for cadmium analysis
title_full Development of polyaniline microarray electrodes for cadmium analysis
title_fullStr Development of polyaniline microarray electrodes for cadmium analysis
title_full_unstemmed Development of polyaniline microarray electrodes for cadmium analysis
title_sort Development of polyaniline microarray electrodes for cadmium analysis
author Oliveira, Marta
author_facet Oliveira, Marta
Viswanathan, Subramanian
Morais, Simone
Delerue-Matos, Cristina
author_role author
author2 Viswanathan, Subramanian
Morais, Simone
Delerue-Matos, Cristina
author2_role author
author
author
dc.contributor.none.fl_str_mv REPOSITÓRIO P.PORTO
dc.contributor.author.fl_str_mv Oliveira, Marta
Viswanathan, Subramanian
Morais, Simone
Delerue-Matos, Cristina
dc.subject.por.fl_str_mv Screen-printed electrodes
Microarray electrodes
Aniline
Cadmium
topic Screen-printed electrodes
Microarray electrodes
Aniline
Cadmium
description Disposable screen-printed electrodes (SPCE) were modified using a cosmetic product to partially block the electrode surface in order to obtain a microelectrode array. The microarrays formed were electropolymerized with aniline. Scanning electron microscopy was used to evaluate the modified and polymerized electrode surface. Electrochemical characteristics of the constructed sensor for cadmium analysis were evaluated by cyclic and square-wave voltammetry. Optimized stripping procedure in which the preconcentration of cadmium was achieved by depositing at –1.20 V (vs. Ag/AgCl) resulted in a well defined anodic peak at approximately –0.7 V at pH 4.6. The achieved limit of detection was 4 × 10−9 mol dm−3. Spray modified and polymerized microarray electrodes were successfully applied to quantify cadmium in fish sample digests.
publishDate 2012
dc.date.none.fl_str_mv 2012
2012-01-01T00:00:00Z
2013-10-01T09:29:18Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.22/2043
url http://hdl.handle.net/10400.22/2043
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 1336-9075
10.2478/s11696-012-0175-x
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Springer Verlag
publisher.none.fl_str_mv Springer Verlag
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833600531953614848