Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging

Bibliographic Details
Main Author: António, Débora de Souza
Publication Date: 2018
Format: Master thesis
Language: eng
Source: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Download full: http://hdl.handle.net/10362/59605
Summary: Cancer is the second leading cause of death in the world, and therefore, there is an undeniable need to ensure early screening and detection systems worldwide. The aim of this project is to study the feasibility of a Cone Beam Computed Tomography (CBCT) scanner for simultaneous breast and lung imaging. Additionally, the development of reconstruction algorithms and the study of their impact to the image quality was considered. Monte Carlo (MC) simulations were performed using the PENELOPE code system. A MC geometry model of a CBCT scanner was implemented for energies of 30 keV and 80 keV for hypothetical scanning protocols. Microcalcifications were inserted into the breast and lung of the computational phantom (ICRP Adult Female Reference), used in the simulations for dose assessment and projection acquisition. Reconstructed images were analyzed in terms of the Contrast-to-Noise Ratio (CNR) and dose calculations were performed for two protocols, one with a normalization factor of 2 mGy in the breast and another with 5 mGy in the lungs. Both, MC geometry model and reconstruction algorithm were validated by means of on-field measurements and data acquisition in a clinical center. Dosimetric and imaging performances were evaluated through Quality Assurance phantoms (Computed Tomography Dose Index and Catphan, respectively). Results indicate that the best implementation of the reconstruction algorithm was achieved with 80 keV, using the Hanning filter and linear interpolation. More specifically, for a spherical lung lesion with a radius of 7 mm a 30% CNR gain was found when the number of projections varied from 12 to 36 (corresponding to a dose increase of a factor of 3). This research suggests the possibility of developing a CBCT modulated beam scanner for simultaneous breast and lung imaging while ensuring dose reduction. However further investigation regarding the number of projections needed for image reconstruction is required.
id RCAP_4089ad77657f7f10727f72b94ffc8ddf
oai_identifier_str oai:run.unl.pt:10362/59605
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imagingCone Beam Computed TomographyCancer DetectionMonte Carlo SimulationOrgan DoseMedical Image ReconstructionLung and Breast ImagingDomínio/Área Científica::Engenharia e Tecnologia::Outras Engenharias e TecnologiasCancer is the second leading cause of death in the world, and therefore, there is an undeniable need to ensure early screening and detection systems worldwide. The aim of this project is to study the feasibility of a Cone Beam Computed Tomography (CBCT) scanner for simultaneous breast and lung imaging. Additionally, the development of reconstruction algorithms and the study of their impact to the image quality was considered. Monte Carlo (MC) simulations were performed using the PENELOPE code system. A MC geometry model of a CBCT scanner was implemented for energies of 30 keV and 80 keV for hypothetical scanning protocols. Microcalcifications were inserted into the breast and lung of the computational phantom (ICRP Adult Female Reference), used in the simulations for dose assessment and projection acquisition. Reconstructed images were analyzed in terms of the Contrast-to-Noise Ratio (CNR) and dose calculations were performed for two protocols, one with a normalization factor of 2 mGy in the breast and another with 5 mGy in the lungs. Both, MC geometry model and reconstruction algorithm were validated by means of on-field measurements and data acquisition in a clinical center. Dosimetric and imaging performances were evaluated through Quality Assurance phantoms (Computed Tomography Dose Index and Catphan, respectively). Results indicate that the best implementation of the reconstruction algorithm was achieved with 80 keV, using the Hanning filter and linear interpolation. More specifically, for a spherical lung lesion with a radius of 7 mm a 30% CNR gain was found when the number of projections varied from 12 to 36 (corresponding to a dose increase of a factor of 3). This research suggests the possibility of developing a CBCT modulated beam scanner for simultaneous breast and lung imaging while ensuring dose reduction. However further investigation regarding the number of projections needed for image reconstruction is required.Di Maria, SalvatoreMatela, NunoVaz, JoséRUNAntónio, Débora de Souza2019-02-05T13:55:10Z2018-1220182018-12-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/masterThesisapplication/pdfhttp://hdl.handle.net/10362/59605enginfo:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2024-05-22T17:37:02Zoai:run.unl.pt:10362/59605Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-28T17:08:06.906082Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
title Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
spellingShingle Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
António, Débora de Souza
Cone Beam Computed Tomography
Cancer Detection
Monte Carlo Simulation
Organ Dose
Medical Image Reconstruction
Lung and Breast Imaging
Domínio/Área Científica::Engenharia e Tecnologia::Outras Engenharias e Tecnologias
title_short Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
title_full Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
title_fullStr Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
title_full_unstemmed Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
title_sort Dose assessment and reconstruction algorithm optimization in simultaneous breast and lung CT imaging
author António, Débora de Souza
author_facet António, Débora de Souza
author_role author
dc.contributor.none.fl_str_mv Di Maria, Salvatore
Matela, Nuno
Vaz, José
RUN
dc.contributor.author.fl_str_mv António, Débora de Souza
dc.subject.por.fl_str_mv Cone Beam Computed Tomography
Cancer Detection
Monte Carlo Simulation
Organ Dose
Medical Image Reconstruction
Lung and Breast Imaging
Domínio/Área Científica::Engenharia e Tecnologia::Outras Engenharias e Tecnologias
topic Cone Beam Computed Tomography
Cancer Detection
Monte Carlo Simulation
Organ Dose
Medical Image Reconstruction
Lung and Breast Imaging
Domínio/Área Científica::Engenharia e Tecnologia::Outras Engenharias e Tecnologias
description Cancer is the second leading cause of death in the world, and therefore, there is an undeniable need to ensure early screening and detection systems worldwide. The aim of this project is to study the feasibility of a Cone Beam Computed Tomography (CBCT) scanner for simultaneous breast and lung imaging. Additionally, the development of reconstruction algorithms and the study of their impact to the image quality was considered. Monte Carlo (MC) simulations were performed using the PENELOPE code system. A MC geometry model of a CBCT scanner was implemented for energies of 30 keV and 80 keV for hypothetical scanning protocols. Microcalcifications were inserted into the breast and lung of the computational phantom (ICRP Adult Female Reference), used in the simulations for dose assessment and projection acquisition. Reconstructed images were analyzed in terms of the Contrast-to-Noise Ratio (CNR) and dose calculations were performed for two protocols, one with a normalization factor of 2 mGy in the breast and another with 5 mGy in the lungs. Both, MC geometry model and reconstruction algorithm were validated by means of on-field measurements and data acquisition in a clinical center. Dosimetric and imaging performances were evaluated through Quality Assurance phantoms (Computed Tomography Dose Index and Catphan, respectively). Results indicate that the best implementation of the reconstruction algorithm was achieved with 80 keV, using the Hanning filter and linear interpolation. More specifically, for a spherical lung lesion with a radius of 7 mm a 30% CNR gain was found when the number of projections varied from 12 to 36 (corresponding to a dose increase of a factor of 3). This research suggests the possibility of developing a CBCT modulated beam scanner for simultaneous breast and lung imaging while ensuring dose reduction. However further investigation regarding the number of projections needed for image reconstruction is required.
publishDate 2018
dc.date.none.fl_str_mv 2018-12
2018
2018-12-01T00:00:00Z
2019-02-05T13:55:10Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/masterThesis
format masterThesis
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10362/59605
url http://hdl.handle.net/10362/59605
dc.language.iso.fl_str_mv eng
language eng
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833596459639898112