Are the IR cameras FLIR ONE suitable for clinical applications?

Bibliographic Details
Main Author: Ricardo Vardasca, PhD, ASIS, FRPS
Publication Date: 2019
Format: Article
Language: eng
Source: Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
Download full: http://hdl.handle.net/10400.26/38561
Summary: BACKGROUND: Infrared Thermography (IRT) has been used in clinical environments for at least six decades. In 2014 affordable and attractive low-cost infrared cameras were introduced into the market that facilitated the ability of being at- tached to mobile devices such as tablets and smartphones. Despite these cameras do not satisfy the minimum specifica- tions recommended for clinical use they have already been used in clinical applications. It is therefore important to verify the performance of these devices. In this paper the start-up stability and the absolute temperature offset, in particular within the temperature range of the human body, are evaluated using the android and iOS OEM connection versions of the FLIR ONE IR 2nd generation and compared from the end user point of view. MATERIALS AND METHODS: Four FLIR ONE IR 2nd generation cameras were used, two developed to be attached to android systems and the other two to apple iOS systems. A start-up drift test at 30 ºC and a temperature sweep from 20 ºC to 40 ºC in steps of 1 ºC, representing the human body temperature range, were carried out. For the temperature perfor- mance assessments, a blackbody Isotech Hyperion R Model 982 was used as temperature reference (uncertainty of ±0.1°C). It was first set at 30ºC temperature reference for the start-up drift test, the cameras along with the attached de- vices were switched on and measurements were taken at five-minute intervals for one hour at a distance of 30 cm from the blackbody target. For the temperature sweep, the blackbody reference was set to 20ºC. Images were then taken with the IRT imaging devices and readings were taken while the blackbody setpoint was increased in steps of 1ºC up to 40ºC, wait- ing 15 minutes in between each step for blackbody temperature stability. RESULTS: The FLIR One 2nd generation thermal cameras overestimate the temperature reading during the start-up off set drift test and take around 15 to 20 minutes to reach measurement stability with an average difference from the blackbody indicated temperature of 0.9 ºC. In the human body temperature offset drift test there is a higher difference from the calibration source at temperatures below 24 ºC, where the temperature readouts are more than 2.0 °C above the real temperatures set at the blackbody. There is a high interclass correlation between the thermal cameras' readings and the calibration source set temperatures and also between the measurements of the two OEM versions of the thermal cam eras studied. The span of limits of agreement (LOA) of the measurements of all FLIR ONE 2nd generation cameras with the blackbody temperature was 2.23 ºC. CONCLUSIONS: Despite these systems being attractive in price and manufacturer provided features, their operational performance does not comply with the required standards for clinical use. The thermal information provided by these im aging systems should only be taken into account for monitoring purposes, as some previous research demonstrated, and not as an input for diagnostic judgments, if they require absolute temperature values to be correct. It is important to note that the cameras provider does not advertise them as medical devices.
id RCAP_1e1559aa51c5e794f76564cbf244df1c
oai_identifier_str oai:comum.rcaap.pt:10400.26/38561
network_acronym_str RCAP
network_name_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository_id_str https://opendoar.ac.uk/repository/7160
spelling Are the IR cameras FLIR ONE suitable for clinical applications?Affordable IR camerasFLIR ONE 2nd genInfrared thermographymobile thermographyBACKGROUND: Infrared Thermography (IRT) has been used in clinical environments for at least six decades. In 2014 affordable and attractive low-cost infrared cameras were introduced into the market that facilitated the ability of being at- tached to mobile devices such as tablets and smartphones. Despite these cameras do not satisfy the minimum specifica- tions recommended for clinical use they have already been used in clinical applications. It is therefore important to verify the performance of these devices. In this paper the start-up stability and the absolute temperature offset, in particular within the temperature range of the human body, are evaluated using the android and iOS OEM connection versions of the FLIR ONE IR 2nd generation and compared from the end user point of view. MATERIALS AND METHODS: Four FLIR ONE IR 2nd generation cameras were used, two developed to be attached to android systems and the other two to apple iOS systems. A start-up drift test at 30 ºC and a temperature sweep from 20 ºC to 40 ºC in steps of 1 ºC, representing the human body temperature range, were carried out. For the temperature perfor- mance assessments, a blackbody Isotech Hyperion R Model 982 was used as temperature reference (uncertainty of ±0.1°C). It was first set at 30ºC temperature reference for the start-up drift test, the cameras along with the attached de- vices were switched on and measurements were taken at five-minute intervals for one hour at a distance of 30 cm from the blackbody target. For the temperature sweep, the blackbody reference was set to 20ºC. Images were then taken with the IRT imaging devices and readings were taken while the blackbody setpoint was increased in steps of 1ºC up to 40ºC, wait- ing 15 minutes in between each step for blackbody temperature stability. RESULTS: The FLIR One 2nd generation thermal cameras overestimate the temperature reading during the start-up off set drift test and take around 15 to 20 minutes to reach measurement stability with an average difference from the blackbody indicated temperature of 0.9 ºC. In the human body temperature offset drift test there is a higher difference from the calibration source at temperatures below 24 ºC, where the temperature readouts are more than 2.0 °C above the real temperatures set at the blackbody. There is a high interclass correlation between the thermal cameras' readings and the calibration source set temperatures and also between the measurements of the two OEM versions of the thermal cam eras studied. The span of limits of agreement (LOA) of the measurements of all FLIR ONE 2nd generation cameras with the blackbody temperature was 2.23 ºC. CONCLUSIONS: Despite these systems being attractive in price and manufacturer provided features, their operational performance does not comply with the required standards for clinical use. The thermal information provided by these im aging systems should only be taken into account for monitoring purposes, as some previous research demonstrated, and not as an input for diagnostic judgments, if they require absolute temperature values to be correct. It is important to note that the cameras provider does not advertise them as medical devices.Repositório ComumRicardo Vardasca, PhD, ASIS, FRPS2022-01-04T10:46:46Z20192019-01-01T00:00:00Zinfo:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/articleapplication/pdfhttp://hdl.handle.net/10400.26/38561enginfo:eu-repo/semantics/openAccessreponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiainstacron:RCAAP2025-05-05T14:38:01Zoai:comum.rcaap.pt:10400.26/38561Portal AgregadorONGhttps://www.rcaap.pt/oai/openaireinfo@rcaap.ptopendoar:https://opendoar.ac.uk/repository/71602025-05-29T07:02:07.985808Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologiafalse
dc.title.none.fl_str_mv Are the IR cameras FLIR ONE suitable for clinical applications?
title Are the IR cameras FLIR ONE suitable for clinical applications?
spellingShingle Are the IR cameras FLIR ONE suitable for clinical applications?
Ricardo Vardasca, PhD, ASIS, FRPS
Affordable IR cameras
FLIR ONE 2nd gen
Infrared thermography
mobile thermography
title_short Are the IR cameras FLIR ONE suitable for clinical applications?
title_full Are the IR cameras FLIR ONE suitable for clinical applications?
title_fullStr Are the IR cameras FLIR ONE suitable for clinical applications?
title_full_unstemmed Are the IR cameras FLIR ONE suitable for clinical applications?
title_sort Are the IR cameras FLIR ONE suitable for clinical applications?
author Ricardo Vardasca, PhD, ASIS, FRPS
author_facet Ricardo Vardasca, PhD, ASIS, FRPS
author_role author
dc.contributor.none.fl_str_mv Repositório Comum
dc.contributor.author.fl_str_mv Ricardo Vardasca, PhD, ASIS, FRPS
dc.subject.por.fl_str_mv Affordable IR cameras
FLIR ONE 2nd gen
Infrared thermography
mobile thermography
topic Affordable IR cameras
FLIR ONE 2nd gen
Infrared thermography
mobile thermography
description BACKGROUND: Infrared Thermography (IRT) has been used in clinical environments for at least six decades. In 2014 affordable and attractive low-cost infrared cameras were introduced into the market that facilitated the ability of being at- tached to mobile devices such as tablets and smartphones. Despite these cameras do not satisfy the minimum specifica- tions recommended for clinical use they have already been used in clinical applications. It is therefore important to verify the performance of these devices. In this paper the start-up stability and the absolute temperature offset, in particular within the temperature range of the human body, are evaluated using the android and iOS OEM connection versions of the FLIR ONE IR 2nd generation and compared from the end user point of view. MATERIALS AND METHODS: Four FLIR ONE IR 2nd generation cameras were used, two developed to be attached to android systems and the other two to apple iOS systems. A start-up drift test at 30 ºC and a temperature sweep from 20 ºC to 40 ºC in steps of 1 ºC, representing the human body temperature range, were carried out. For the temperature perfor- mance assessments, a blackbody Isotech Hyperion R Model 982 was used as temperature reference (uncertainty of ±0.1°C). It was first set at 30ºC temperature reference for the start-up drift test, the cameras along with the attached de- vices were switched on and measurements were taken at five-minute intervals for one hour at a distance of 30 cm from the blackbody target. For the temperature sweep, the blackbody reference was set to 20ºC. Images were then taken with the IRT imaging devices and readings were taken while the blackbody setpoint was increased in steps of 1ºC up to 40ºC, wait- ing 15 minutes in between each step for blackbody temperature stability. RESULTS: The FLIR One 2nd generation thermal cameras overestimate the temperature reading during the start-up off set drift test and take around 15 to 20 minutes to reach measurement stability with an average difference from the blackbody indicated temperature of 0.9 ºC. In the human body temperature offset drift test there is a higher difference from the calibration source at temperatures below 24 ºC, where the temperature readouts are more than 2.0 °C above the real temperatures set at the blackbody. There is a high interclass correlation between the thermal cameras' readings and the calibration source set temperatures and also between the measurements of the two OEM versions of the thermal cam eras studied. The span of limits of agreement (LOA) of the measurements of all FLIR ONE 2nd generation cameras with the blackbody temperature was 2.23 ºC. CONCLUSIONS: Despite these systems being attractive in price and manufacturer provided features, their operational performance does not comply with the required standards for clinical use. The thermal information provided by these im aging systems should only be taken into account for monitoring purposes, as some previous research demonstrated, and not as an input for diagnostic judgments, if they require absolute temperature values to be correct. It is important to note that the cameras provider does not advertise them as medical devices.
publishDate 2019
dc.date.none.fl_str_mv 2019
2019-01-01T00:00:00Z
2022-01-04T10:46:46Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://hdl.handle.net/10400.26/38561
url http://hdl.handle.net/10400.26/38561
dc.language.iso.fl_str_mv eng
language eng
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.source.none.fl_str_mv reponame:Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
instname:FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron:RCAAP
instname_str FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
instacron_str RCAAP
institution RCAAP
reponame_str Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
collection Repositórios Científicos de Acesso Aberto de Portugal (RCAAP)
repository.name.fl_str_mv Repositórios Científicos de Acesso Aberto de Portugal (RCAAP) - FCCN, serviços digitais da FCT – Fundação para a Ciência e a Tecnologia
repository.mail.fl_str_mv info@rcaap.pt
_version_ 1833602865645486080