Microperimetry differences in macular sensitivity threshold between first and second tested eyes

Bibliographic Details
Main Author: Junqueira,Natacha B.
Publication Date: 2021
Other Authors: Lima,Luiz H., Ferreira,Rodrigo B., Garcia,Denny Marcos, Furtado,João M., Jorge,Rodrigo
Format: Article
Language: eng
Source: Arquivos brasileiros de oftalmologia (Online)
Download full: http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203
Summary: ABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test.
id CBO-2_0062ad50552e4f17f111b1a625e17dae
oai_identifier_str oai:scielo:S0004-27492021000300203
network_acronym_str CBO-2
network_name_str Arquivos brasileiros de oftalmologia (Online)
repository_id_str
spelling Microperimetry differences in macular sensitivity threshold between first and second tested eyesMacula luteaFixation, ocularBiasVisual field testsVisual acuityABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test.Conselho Brasileiro de Oftalmologia2021-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203Arquivos Brasileiros de Oftalmologia v.84 n.3 2021reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.5935/0004-2749.20210034info:eu-repo/semantics/openAccessJunqueira,Natacha B.Lima,Luiz H.Ferreira,Rodrigo B.Garcia,Denny MarcosFurtado,João M.Jorge,Rodrigoeng2021-05-18T00:00:00Zoai:scielo:S0004-27492021000300203Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2021-05-18T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false
dc.title.none.fl_str_mv Microperimetry differences in macular sensitivity threshold between first and second tested eyes
title Microperimetry differences in macular sensitivity threshold between first and second tested eyes
spellingShingle Microperimetry differences in macular sensitivity threshold between first and second tested eyes
Junqueira,Natacha B.
Macula lutea
Fixation, ocular
Bias
Visual field tests
Visual acuity
title_short Microperimetry differences in macular sensitivity threshold between first and second tested eyes
title_full Microperimetry differences in macular sensitivity threshold between first and second tested eyes
title_fullStr Microperimetry differences in macular sensitivity threshold between first and second tested eyes
title_full_unstemmed Microperimetry differences in macular sensitivity threshold between first and second tested eyes
title_sort Microperimetry differences in macular sensitivity threshold between first and second tested eyes
author Junqueira,Natacha B.
author_facet Junqueira,Natacha B.
Lima,Luiz H.
Ferreira,Rodrigo B.
Garcia,Denny Marcos
Furtado,João M.
Jorge,Rodrigo
author_role author
author2 Lima,Luiz H.
Ferreira,Rodrigo B.
Garcia,Denny Marcos
Furtado,João M.
Jorge,Rodrigo
author2_role author
author
author
author
author
dc.contributor.author.fl_str_mv Junqueira,Natacha B.
Lima,Luiz H.
Ferreira,Rodrigo B.
Garcia,Denny Marcos
Furtado,João M.
Jorge,Rodrigo
dc.subject.por.fl_str_mv Macula lutea
Fixation, ocular
Bias
Visual field tests
Visual acuity
topic Macula lutea
Fixation, ocular
Bias
Visual field tests
Visual acuity
description ABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test.
publishDate 2021
dc.date.none.fl_str_mv 2021-06-01
dc.type.driver.fl_str_mv info:eu-repo/semantics/article
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
format article
status_str publishedVersion
dc.identifier.uri.fl_str_mv http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203
url http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203
dc.language.iso.fl_str_mv eng
language eng
dc.relation.none.fl_str_mv 10.5935/0004-2749.20210034
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv text/html
dc.publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
publisher.none.fl_str_mv Conselho Brasileiro de Oftalmologia
dc.source.none.fl_str_mv Arquivos Brasileiros de Oftalmologia v.84 n.3 2021
reponame:Arquivos brasileiros de oftalmologia (Online)
instname:Conselho Brasileiro de Oftalmologia (CBO)
instacron:CBO
instname_str Conselho Brasileiro de Oftalmologia (CBO)
instacron_str CBO
institution CBO
reponame_str Arquivos brasileiros de oftalmologia (Online)
collection Arquivos brasileiros de oftalmologia (Online)
repository.name.fl_str_mv Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)
repository.mail.fl_str_mv aboonline@cbo.com.br||abo@cbo.com.br
_version_ 1754209031327580160