Microperimetry differences in macular sensitivity threshold between first and second tested eyes
Main Author: | |
---|---|
Publication Date: | 2021 |
Other Authors: | , , , , |
Format: | Article |
Language: | eng |
Source: | Arquivos brasileiros de oftalmologia (Online) |
Download full: | http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203 |
Summary: | ABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test. |
id |
CBO-2_0062ad50552e4f17f111b1a625e17dae |
---|---|
oai_identifier_str |
oai:scielo:S0004-27492021000300203 |
network_acronym_str |
CBO-2 |
network_name_str |
Arquivos brasileiros de oftalmologia (Online) |
repository_id_str |
|
spelling |
Microperimetry differences in macular sensitivity threshold between first and second tested eyesMacula luteaFixation, ocularBiasVisual field testsVisual acuityABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test.Conselho Brasileiro de Oftalmologia2021-06-01info:eu-repo/semantics/articleinfo:eu-repo/semantics/publishedVersiontext/htmlhttp://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203Arquivos Brasileiros de Oftalmologia v.84 n.3 2021reponame:Arquivos brasileiros de oftalmologia (Online)instname:Conselho Brasileiro de Oftalmologia (CBO)instacron:CBO10.5935/0004-2749.20210034info:eu-repo/semantics/openAccessJunqueira,Natacha B.Lima,Luiz H.Ferreira,Rodrigo B.Garcia,Denny MarcosFurtado,João M.Jorge,Rodrigoeng2021-05-18T00:00:00Zoai:scielo:S0004-27492021000300203Revistahttp://aboonline.org.br/https://old.scielo.br/oai/scielo-oai.phpaboonline@cbo.com.br||abo@cbo.com.br1678-29250004-2749opendoar:2021-05-18T00:00Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO)false |
dc.title.none.fl_str_mv |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
title |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
spellingShingle |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes Junqueira,Natacha B. Macula lutea Fixation, ocular Bias Visual field tests Visual acuity |
title_short |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
title_full |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
title_fullStr |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
title_full_unstemmed |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
title_sort |
Microperimetry differences in macular sensitivity threshold between first and second tested eyes |
author |
Junqueira,Natacha B. |
author_facet |
Junqueira,Natacha B. Lima,Luiz H. Ferreira,Rodrigo B. Garcia,Denny Marcos Furtado,João M. Jorge,Rodrigo |
author_role |
author |
author2 |
Lima,Luiz H. Ferreira,Rodrigo B. Garcia,Denny Marcos Furtado,João M. Jorge,Rodrigo |
author2_role |
author author author author author |
dc.contributor.author.fl_str_mv |
Junqueira,Natacha B. Lima,Luiz H. Ferreira,Rodrigo B. Garcia,Denny Marcos Furtado,João M. Jorge,Rodrigo |
dc.subject.por.fl_str_mv |
Macula lutea Fixation, ocular Bias Visual field tests Visual acuity |
topic |
Macula lutea Fixation, ocular Bias Visual field tests Visual acuity |
description |
ABSTRACT Purpose: To comparatively assess the macular sensitivity threshold of microperimetry and the fixation stability between the first (right) and second (left) tested eye of normal participants. Methods: Thirty healthy patients were randomly assigned to two groups. The participants underwent microperimetry in the fast mode and expert mode in groups I and II, respectively. Each participant underwent a single test and the right eye was tested first. Results: The mean macular sensitivity threshold (± standard deviation [SD]) was 24.5 ± 2.3 dB and 25.7 ± 1.1 dB in the first (right) and second (left) eyes of group I, respectively (p=0.0415) and 26.7 ± 4.5 dB and 27.3 ± 4.0 dB in the first (right) and second (left) eyes of group II, respectively (p=0.58). There was no statistically significant difference between eyes in either group (p=0.1512). Regarding fixation stability (evaluated in the microperimetry expert mode group), the mean ± SD percentage of fixation points within the 1-degree central macula (P1) was 87.9 ± 11.5% in the right eye and 93.8 ± 6.6% in the left eye. The paired t-test did not show a statistically significant difference between eyes (p=0.140). Mean ± SD P2 value was 95.5 ± 4.9% in the right eye and 98.5 ± 2.1% in the left eye. The analysis demonstrated an increase in the percentage of fixation points in the second tested eye compared with the first one (paired t-test= 2.364; p=0.034). There was a negative correlation between the macular sensitivity threshold of the right eye and the duration of the examination for both groups (microperimetry expert mode: r=-0.717; p=0.0026; microperimetry in the fast mode: r=-0.843; p<0.0001). Conclusion: Mean macular sensitivity threshold was higher in the second tested eye in the microperimetry in the fast mode group and was similar in both eyes in the expert mode. Our data suggest that comprehension of the examination by the individual may impact the results of the microperimetry test. |
publishDate |
2021 |
dc.date.none.fl_str_mv |
2021-06-01 |
dc.type.driver.fl_str_mv |
info:eu-repo/semantics/article |
dc.type.status.fl_str_mv |
info:eu-repo/semantics/publishedVersion |
format |
article |
status_str |
publishedVersion |
dc.identifier.uri.fl_str_mv |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203 |
url |
http://old.scielo.br/scielo.php?script=sci_arttext&pid=S0004-27492021000300203 |
dc.language.iso.fl_str_mv |
eng |
language |
eng |
dc.relation.none.fl_str_mv |
10.5935/0004-2749.20210034 |
dc.rights.driver.fl_str_mv |
info:eu-repo/semantics/openAccess |
eu_rights_str_mv |
openAccess |
dc.format.none.fl_str_mv |
text/html |
dc.publisher.none.fl_str_mv |
Conselho Brasileiro de Oftalmologia |
publisher.none.fl_str_mv |
Conselho Brasileiro de Oftalmologia |
dc.source.none.fl_str_mv |
Arquivos Brasileiros de Oftalmologia v.84 n.3 2021 reponame:Arquivos brasileiros de oftalmologia (Online) instname:Conselho Brasileiro de Oftalmologia (CBO) instacron:CBO |
instname_str |
Conselho Brasileiro de Oftalmologia (CBO) |
instacron_str |
CBO |
institution |
CBO |
reponame_str |
Arquivos brasileiros de oftalmologia (Online) |
collection |
Arquivos brasileiros de oftalmologia (Online) |
repository.name.fl_str_mv |
Arquivos brasileiros de oftalmologia (Online) - Conselho Brasileiro de Oftalmologia (CBO) |
repository.mail.fl_str_mv |
aboonline@cbo.com.br||abo@cbo.com.br |
_version_ |
1754209031327580160 |