Determinação in situ de analitos de interesse alimentício empregando tratamento de imagens digitais de spot tests

Detalhes bibliográficos
Ano de defesa: 2013
Autor(a) principal: Benedetti, Luzia Pires dos Santos
Orientador(a): Fatibello Filho, Orlando lattes
Banca de defesa: Não Informado pela instituição
Tipo de documento: Dissertação
Tipo de acesso: Acesso aberto
Idioma: por
Instituição de defesa: Universidade Federal de São Carlos
Programa de Pós-Graduação: Programa de Pós-Graduação em Química - PPGQ
Departamento: Não Informado pela instituição
País: BR
Palavras-chave em Português:
RGB
Área do conhecimento CNPq:
Link de acesso: https://repositorio.ufscar.br/handle/20.500.14289/6565
Resumo: In this work a simple, fast and inexpensive analytical method was developed to carry out in situ determinations of ethanol and sulfite ions in food samples. The method was based on digital images captured by cameras and / or mobile phones, using a proper device for this purpose. Indirect colorimetry was employed, where it was measured the intensity of the light reflected by the solution, employing the RGB color model. The quantification of the analytes was performed by calibration curve method, using image processing software to obtain the numerical values for each one of the three primary colors (red (R), green (G) and blue (B)). In the first procedure, it was determined the ethanol concentration in drinks, obtaining limits of detection (LOD) and quantification (LOQ) of 0.25 and 0.85 %, respectively, with a linear calibration curve in the range from 1 to 20 % v / v (r = 0.999), using the R channel due to better linearity and results in close agreement to the spectrophotometric method used as reference method, with a confidence level of 95 %. The recovery study was carried out in six commercial samples and the recovery percentage ranged between 85.0 and 106%. The sulfite ions were determined in wines, vinegars and juices samples, with limits of detection (LOD) and quantification (LOQ) of 3.3 × 10-5 mol L -1 and 1.0 × 10 -4 mol L-1, respectively, and with a linear calibration curve from 1.0 × 10-4 to 3.0 × 10-3 mol L1 (r = 0.997) using the B channel supplying a good linearity and results in close agreement to the iodometric titration method employed as reference method, at a confidence level of 95 %. The recovery percentage of sulfite in commercial samples ranged between 84.4 and 103%.