Avaliação de doses ocupacionais em cirurgias ortopédicas

Bibliographic Details
Main Author: Endlich, Larissa Lopes
Publication Date: 2018
Format: Bachelor thesis
Language: por
Source: Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
Download full: http://repositorio.utfpr.edu.br/jspui/handle/1/17843
Summary: The objective of the present study was measured as doses of occupational radiation at the level of the physician who performs orthopedic surgeries using a fluoroscopy device was exposed in his workday by means of a thermoluminescent dosimetry (TLD), comparing the occupational doses obtained with legislation and propose measures of radioprotection. From the monitoring carried out, after approval of the research by energy committees, between 18/12/2017 to 05/02/2018, reponses were obtained for the targetcritical points: whole body (thorax), crystalline, pneuide, right and left hands. From the average measurement values obtained for a procedure, an annual radiation dose is estimated in each of the analyzed regions and compared with a current legislation. Compared with CNEN 3.01, the effective dose estimate for the entire body (measured in the chest region) above the apron was 109 ± 86 mSv/year, four times higher than the national legislation (20,0 mSv / year), however, if the professional is considered to wear a plumbing apron (IPE) and even reduces a dose by 99,94%, an estimated dose decreased to 0,07 ± 0,05 mSv/year, being appropriate. As estimates of equivalent doses for the extremities (79 ± 55 mSv/year) and thyroid (61 ± 99 mSv/year), results are obtained under the current legislation: 500 mSv/year and 300 mSv / year for extremities and thyroid, respectively. (20 mSv/year): 79 ± 64 mSv/year. The results confirm the need to comply with the national laws in force through a monitoring program of the professional who is exposed to ionizing radiation, in addition to the use of radiation protection clothing such as apron, tire protector, gloves and goggles; focusing on the solo medical doctor, but on a whole team involved in the surgery.
id UTFPR-12_39f8c0abac38c5aa0a3490b211f73a5a
oai_identifier_str oai:repositorio.utfpr.edu.br:1/17843
network_acronym_str UTFPR-12
network_name_str Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
repository_id_str
spelling Avaliação de doses ocupacionais em cirurgias ortopédicasEvaluation of occupational doses in orthopedic surgeriesCirurgia ortopédicaFluoroscopiaRadiação - DosimetriaOrthopedic surgeryFluoroscopyRadiation dosimetryCNPQ::ENGENHARIAS::ENGENHARIA DE PRODUCAO::GERENCIA DE PRODUCAO::HIGIENE E SEGURANCA DO TRABALHOThe objective of the present study was measured as doses of occupational radiation at the level of the physician who performs orthopedic surgeries using a fluoroscopy device was exposed in his workday by means of a thermoluminescent dosimetry (TLD), comparing the occupational doses obtained with legislation and propose measures of radioprotection. From the monitoring carried out, after approval of the research by energy committees, between 18/12/2017 to 05/02/2018, reponses were obtained for the targetcritical points: whole body (thorax), crystalline, pneuide, right and left hands. From the average measurement values obtained for a procedure, an annual radiation dose is estimated in each of the analyzed regions and compared with a current legislation. Compared with CNEN 3.01, the effective dose estimate for the entire body (measured in the chest region) above the apron was 109 ± 86 mSv/year, four times higher than the national legislation (20,0 mSv / year), however, if the professional is considered to wear a plumbing apron (IPE) and even reduces a dose by 99,94%, an estimated dose decreased to 0,07 ± 0,05 mSv/year, being appropriate. As estimates of equivalent doses for the extremities (79 ± 55 mSv/year) and thyroid (61 ± 99 mSv/year), results are obtained under the current legislation: 500 mSv/year and 300 mSv / year for extremities and thyroid, respectively. (20 mSv/year): 79 ± 64 mSv/year. The results confirm the need to comply with the national laws in force through a monitoring program of the professional who is exposed to ionizing radiation, in addition to the use of radiation protection clothing such as apron, tire protector, gloves and goggles; focusing on the solo medical doctor, but on a whole team involved in the surgery.O objetivo do presente trabalho foi medir as doses de radiação ocupacionais ao qual um médico que realiza cirurgias ortopédicas utilizando um aparelho de fluoroscopia estava exposto em sua jornada de trabalho por meio de dosimetria do tipo termoluminescente (TLD) realizando uma comparação das doses ocupacionais obtidas com a legislação em vigor e propor medidas de radioproteção. A partir do monitoramento realizado, após a aprovação da pesquisa pelos comitês de ética, entre os dias 18/12/2017 a 05/02/2018, foram obtidas respostas para os pontos alvo-críticos: corpo inteiro (tórax), cristalino, tireoide, mãos direita e esquerda. A partir dos valores de dose médios obtidos para um procedimento, estimou-se a dose de radiação anual em cada uma das regiões analisadas e comparou-se com a legislação vigente. A comparação com a norma CNEN 3.01 demonstrou que a estimativa de dose efetiva para o corpo inteiro (medida na região do tórax) por cima do avental teve resultado (109 ± 86 mSv/ano), cinco vezes mais alto do que a legislação nacional (20,0 mSv/ano), porém, se considerado que o profissional trabalha utilizando um avental plumbífero (EPI) e o mesmo reduz a dose em 99,94%, a dose estimada diminui para 0,07 ± 0,05 mSv/ano, estando adequada. As estimativas de doses equivalentes para as extremidades (79 ± 55 mSv/ano) e tireoide (61 ± 99 mSv/ano) tiveram resultados dentro das legislações vigentes: 500 mSv/ano e 300 mSv/ano para extremidades e tireoide, respectivamente. Entretanto, a dose equivalente para o cristalino (ponto de medida na testa) teve resultado quatro vezes mais alto do que o indicado na legislação (20 mSv/ano): 79 ± 64 mSv/ano. Os resultados confirmam a necessidade do cumprimento das legislações nacionais em vigor através de um programa de monitoramento do profissional que está exposto à radiação ionizante, além do uso de vestimentas de proteção radiológica como avental, protetor de tireoide, luvas e óculos plumbíferos; focando não somente no médico ortopedista, mas em toda a equipe envolvida na cirurgia.Universidade Tecnológica Federal do ParanáCuritibaBrasilEspecialização em Engenharia de Segurança do TrabalhoUTFPRFilipov, DanielleCorrêa, Janine NicolosiCatai, Rodrigo EduardoMatoski, AdalbertoHara, Massayuki MárioEndlich, Larissa Lopes2020-11-20T20:00:58Z2020-11-20T20:00:58Z2018-03-21info:eu-repo/semantics/publishedVersioninfo:eu-repo/semantics/bachelorThesisapplication/pdfENDLICH, Larissa Lopes. Avaliação de doses ocupacionais em cirurgias ortopédicas. 2018. 57 f. Trabalho de Conclusão de Curso (Especialização em Engenharia de Segurança do Trabalho) - Universidade Tecnológica Federal do Paraná, Curitiba, 2018.http://repositorio.utfpr.edu.br/jspui/handle/1/17843porinfo:eu-repo/semantics/openAccessreponame:Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))instname:Universidade Tecnológica Federal do Paraná (UTFPR)instacron:UTFPR2020-11-20T20:00:58Zoai:repositorio.utfpr.edu.br:1/17843Repositório InstitucionalPUBhttp://repositorio.utfpr.edu.br:8080/oai/requestriut@utfpr.edu.bropendoar:2020-11-20T20:00:58Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT)) - Universidade Tecnológica Federal do Paraná (UTFPR)false
dc.title.none.fl_str_mv Avaliação de doses ocupacionais em cirurgias ortopédicas
Evaluation of occupational doses in orthopedic surgeries
title Avaliação de doses ocupacionais em cirurgias ortopédicas
spellingShingle Avaliação de doses ocupacionais em cirurgias ortopédicas
Endlich, Larissa Lopes
Cirurgia ortopédica
Fluoroscopia
Radiação - Dosimetria
Orthopedic surgery
Fluoroscopy
Radiation dosimetry
CNPQ::ENGENHARIAS::ENGENHARIA DE PRODUCAO::GERENCIA DE PRODUCAO::HIGIENE E SEGURANCA DO TRABALHO
title_short Avaliação de doses ocupacionais em cirurgias ortopédicas
title_full Avaliação de doses ocupacionais em cirurgias ortopédicas
title_fullStr Avaliação de doses ocupacionais em cirurgias ortopédicas
title_full_unstemmed Avaliação de doses ocupacionais em cirurgias ortopédicas
title_sort Avaliação de doses ocupacionais em cirurgias ortopédicas
author Endlich, Larissa Lopes
author_facet Endlich, Larissa Lopes
author_role author
dc.contributor.none.fl_str_mv Filipov, Danielle
Corrêa, Janine Nicolosi
Catai, Rodrigo Eduardo
Matoski, Adalberto
Hara, Massayuki Mário
dc.contributor.author.fl_str_mv Endlich, Larissa Lopes
dc.subject.por.fl_str_mv Cirurgia ortopédica
Fluoroscopia
Radiação - Dosimetria
Orthopedic surgery
Fluoroscopy
Radiation dosimetry
CNPQ::ENGENHARIAS::ENGENHARIA DE PRODUCAO::GERENCIA DE PRODUCAO::HIGIENE E SEGURANCA DO TRABALHO
topic Cirurgia ortopédica
Fluoroscopia
Radiação - Dosimetria
Orthopedic surgery
Fluoroscopy
Radiation dosimetry
CNPQ::ENGENHARIAS::ENGENHARIA DE PRODUCAO::GERENCIA DE PRODUCAO::HIGIENE E SEGURANCA DO TRABALHO
description The objective of the present study was measured as doses of occupational radiation at the level of the physician who performs orthopedic surgeries using a fluoroscopy device was exposed in his workday by means of a thermoluminescent dosimetry (TLD), comparing the occupational doses obtained with legislation and propose measures of radioprotection. From the monitoring carried out, after approval of the research by energy committees, between 18/12/2017 to 05/02/2018, reponses were obtained for the targetcritical points: whole body (thorax), crystalline, pneuide, right and left hands. From the average measurement values obtained for a procedure, an annual radiation dose is estimated in each of the analyzed regions and compared with a current legislation. Compared with CNEN 3.01, the effective dose estimate for the entire body (measured in the chest region) above the apron was 109 ± 86 mSv/year, four times higher than the national legislation (20,0 mSv / year), however, if the professional is considered to wear a plumbing apron (IPE) and even reduces a dose by 99,94%, an estimated dose decreased to 0,07 ± 0,05 mSv/year, being appropriate. As estimates of equivalent doses for the extremities (79 ± 55 mSv/year) and thyroid (61 ± 99 mSv/year), results are obtained under the current legislation: 500 mSv/year and 300 mSv / year for extremities and thyroid, respectively. (20 mSv/year): 79 ± 64 mSv/year. The results confirm the need to comply with the national laws in force through a monitoring program of the professional who is exposed to ionizing radiation, in addition to the use of radiation protection clothing such as apron, tire protector, gloves and goggles; focusing on the solo medical doctor, but on a whole team involved in the surgery.
publishDate 2018
dc.date.none.fl_str_mv 2018-03-21
2020-11-20T20:00:58Z
2020-11-20T20:00:58Z
dc.type.status.fl_str_mv info:eu-repo/semantics/publishedVersion
dc.type.driver.fl_str_mv info:eu-repo/semantics/bachelorThesis
format bachelorThesis
status_str publishedVersion
dc.identifier.uri.fl_str_mv ENDLICH, Larissa Lopes. Avaliação de doses ocupacionais em cirurgias ortopédicas. 2018. 57 f. Trabalho de Conclusão de Curso (Especialização em Engenharia de Segurança do Trabalho) - Universidade Tecnológica Federal do Paraná, Curitiba, 2018.
http://repositorio.utfpr.edu.br/jspui/handle/1/17843
identifier_str_mv ENDLICH, Larissa Lopes. Avaliação de doses ocupacionais em cirurgias ortopédicas. 2018. 57 f. Trabalho de Conclusão de Curso (Especialização em Engenharia de Segurança do Trabalho) - Universidade Tecnológica Federal do Paraná, Curitiba, 2018.
url http://repositorio.utfpr.edu.br/jspui/handle/1/17843
dc.language.iso.fl_str_mv por
language por
dc.rights.driver.fl_str_mv info:eu-repo/semantics/openAccess
eu_rights_str_mv openAccess
dc.format.none.fl_str_mv application/pdf
dc.publisher.none.fl_str_mv Universidade Tecnológica Federal do Paraná
Curitiba
Brasil
Especialização em Engenharia de Segurança do Trabalho
UTFPR
publisher.none.fl_str_mv Universidade Tecnológica Federal do Paraná
Curitiba
Brasil
Especialização em Engenharia de Segurança do Trabalho
UTFPR
dc.source.none.fl_str_mv reponame:Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
instname:Universidade Tecnológica Federal do Paraná (UTFPR)
instacron:UTFPR
instname_str Universidade Tecnológica Federal do Paraná (UTFPR)
instacron_str UTFPR
institution UTFPR
reponame_str Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
collection Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT))
repository.name.fl_str_mv Repositório Institucional da UTFPR (da Universidade Tecnológica Federal do Paraná (RIUT)) - Universidade Tecnológica Federal do Paraná (UTFPR)
repository.mail.fl_str_mv riut@utfpr.edu.br
_version_ 1834836308896251904